SOURCES OF ERROR

l1. Setup of the experiment.

The fact that the gap between the different positions where the prove rests is set to 4mm, a smaller range of motion would provide a better reading, specially when finding the stagnation point.

l2. The reference point for the Pilot-Static Tube.

Due to the fact that readings are only taken at distances x/D = 5 and x/D = 10, this calls for a lot of interpolation between data points, and also, a great amount of interpretations and assumptions are to be made based on the overview data given. 

l3. Error in calculating a accurate reading for the equipment.

Due to the fact that the person reading the data had to interpolate from the varying data output from the Voltmeter, this also allows place for error due to averaging.

 

l4. Calculations.

The error due to error propagation in calculations and in the accuracy and reading of the equipment seemed to be minor compared to the error that could occur in the setup of the experiment.  One major source of error could have been the pitot-tube axis not being in alignment with the velocity vector.  Another source of error could have been at the beginning when the center of the cylinder was found to use as a reference point for all further measurements.  The graph of Cp showed one value that seemed to be far from the trend shown by the other points, this is most likely attributed due to a bad pressure reading at the 45-degree pressure port.  The error was calculated using the RMS method and was found to be approximately ±0.5 for the Cd values.  The error in the pressure readings was approximately ±0.69 Pascals.  The only other value used in the velocity calculation was density (no error) and these led to quite small error values in velocity and thus drag force.  From this, we can see that the errors that cannot be mathematically accounted for were more likely to cause problems.